Zhao Jie,Xu Rongkai,Fan Xunchong.Analysis of the influence of different startup conditions on BD test results of pressure steam sterilizer[J].Sterile Supply,2023,02(01):47-50.
Zhao Jie,Xu Rongkai,Fan Xunchong.Analysis of the influence of different startup conditions on BD test results of pressure steam sterilizer[J].Sterile Supply,2023,02(01):47-50. DOI: 10.11910/j.issn.2791-2043.2023.1.07.
Purpose,2,To quantitatively analyze the effect of starting the pressure steam sterilizer in different startup conditions on BD test results, and to provide operational suggestions for medical units.,Methods,2,When the jacket pressure of the pressure steam sterilizer reached the set value before daily sterilization, the BD test program was directly run and set as the control group. When the jacket pressure of the pressure steam sterilizer reached the set value before daily sterilization, the "no-load preheating" program was first run, followed by the BD test program, and set as the experimental group. In the two groups of the sterilization process, 13 sterilization PCDs with increasing difficulty of steam penetration, plus five kinds of chemical indicators and BD test package were placed. After sterilization, record the value of the most difficult PCD and the qualification of the BD test package.,Results,2,The average value of the most difficult PCD passed by the control group was 4.24, and the average value of the most difficult PCD passed by the experimental group was 8.32. The qualified rate of BD in the control group was 89%, and that in the experimental group was 100%; P<0.05, with statistical significance.,Conclusion,2,Preheating without load can improve the sterilization performance of pressure steam sterilizer before BD test every day.
No-load preheatinPCDBDPerformancePressure steam sterilizer
Li Bo-ning, Gu Xin, Wang Yan-hai, et al Cause analysis and countermeasure research on affecting the sterilization quality of the pulsating vacuum pressure steam sterilizer China Medical Equipment 202118(6):146-148.
He Hui-yan, Li Shui-tao, Huang Feng-yu Analysis of influencing factors of sterilization failure caused by non-condensable gas, Chin J Nosocomiol Vol. 31 No. 4 2021.
Industry Standards for the Healthcare Industry of the People’s Republic of China. WS310.3-2016. Part 3 of CSSD of the hospital:surveillance standard for cleaning, disinfection and sterilization [S].2017-06-01.
Wang Fu-wei Influence of pulsation mode of pulsation vacuum sterilizer on BD, PCD test and sterilization effect Medical Equipment 2019,32(12):63-64
Li Wen-yan, Yu Feng-ling Effect of preheating time on BD test of pulsating vacuum sterilizer Journal Of Gan nan Medical University ,2015,35(01):129-130.
EN867-5 Non-biological systems for use in sterilizers-Part 5:Specification for indicator systems and process challenge de-vices for use in performance testing for small sterilizers Type B and Type S[S]. Brussels: European Committee for Standardization
Zhou Juan, Guo Wei, She Xing-wu et al Evaluation of performance of pressure steam sterilizer by lumen sterilization process challenge device with different penetration difficulties Chin J Nosocomiol vol. 30 No.4 2020
National Standard of the People’s Republic of China GB 18282.1-2015 Sterilization of health care products--Chemical indicator Part 1: General requirements 2017-01-01
0
Views
9
Downloads
0
CSCD
Publicity Resources
Related Articles
Related Author
Related Institution