1.Disinfection Supply Center, the First Affiliated Hospital of Soochow University, 215008, Suzhou, Jiangsu, China
Ping Cheng, professor of nursing, research interests: sterilization supply and nursing management, E-mail:chengping197307@suda.edu.cn
Scan for full text
Li Jiayi,Cheng Ping.A practical study on the wet pack incidence of rigid containers applying different structures of silicone protective pads for packaging instruments[J].Sterile Supply,2022,01(03):110-114.
Li Jiayi,Cheng Ping.A practical study on the wet pack incidence of rigid containers applying different structures of silicone protective pads for packaging instruments[J].Sterile Supply,2022,01(03):110-114. DOI: 10.11910/j.issn.2791-2043.2022.3.06.
Objective,2,To study the wet pack incidence of rigid containers applying different structures of silicone protective pads for packaging instruments.,Methods,2,72 surgical instruments were packed in rigid containers and divided into three groups according to the different structures of silicone pads. The rigid container boxes and the instruments loaded in the three groups were the same, and the silicone pads used in groups A and B were flat-bottomed on one side and papillary on the other side. All three groups adopted the same sterilization procedure with pressure steam, and the wet pack incidences was counted after the sterilization procedure.,Results,2,The wet pack incidences in groups A and B was 59.5% and 29.8%, respectively, which was significantly higher than that in group C with 0%. The difference was statistically significant (,P,<,0.05).,Conclusion,2,The wet pack incidence in rigid container boxes with different silicone pads is related to the structure of the built-in silicone pads. In groups A and B, steam is not able to penetrate because of the small density of the silicone pad grid openings. The steam in the box is not smoothly discharged, and condensate accumulation is not easy to dry. In group C, silicone pad with more grids and large grid openings is conducive to sterilization steam penetration, resulting in smooth discharge of the steam in the rigid container box and less condensate accumulation. When the rigid container box applies different structures of silicone protection pads to protect the instruments for packaging instruments, the silicone pads with large grids and flat structures should be adopted to prevent the wet packs and ensure the quality of sterilization.
Rigid containerWet packDifferent structuresSilicone padPressure steam sterilization
ZHU Y H, QIAN Y J. Analysis of application effect of rigid packaging container sterilization box[J]. Chin J Ethnomedicine Ethnopharmacy, 2012, 21(1): 60-61.
DONG M L, LV R L. Selection of optimum program for rigid container in pulsing pre-vacuum pressure steam sterilization[J]. China Med Devices, 2016, 31(12): 110-112.
DONG M L, LV R L. Selection of optimum program for rigid container in pulsing pre-vacuum pressure steam sterilization[J]. China Med Devices, 2016, 31(12): 110-112.
ZHANG J R. Effect of PDCA nursing management in preventing wet package after steam sterilization[J]. China Health Ind, 2019, 16(8): 86-87, 90.
DU S W, YANG Y J, WANG Y J, et al. Effect evaluation of time-interval quality control inspection in CSSD[J]. J Clin Nursing’s Pract, 2017, 2(28): 181, 184.
ZHAO X, WANG L H, ZHAO X, et al. Risk management of healthcare-associated infection in central sterile supply department based on failure mode and effect analysis[J]. Chin J Nosocomiology, 2020, 30(6): 945-950.
HUANG Y, WANG Y F. Effects of different packaging materials and contents on autoclaving wet bags[J]. Chin J Disinfect, 2010, 27(5): 634-635.
Central sterile supply department (CSSD) -- Part 2: Standard for operating procedureof cleaning, disinfection and sterilization (WS310.2-2016). National Health and Family Planning Commission of the People’s Republic of China. 2016:12-27
JING H L, XIAO H, WANG D P. Effect of the ratio of loading capacity to the bottom height of hard container on the incidence of wet bag of knee surgery instruments during steam sterilization[J]. Chin Remedies & Clin, 2020, 20(18): 3167-3168.
JIANG S H. Factor analysis of medical wet packs after sterilization by fishbone diagram and related countermeasures[J]. Nurs Integr Tradit Chin West Med, 2016, 2(10): 133-135.
WANG Li,CHENG Ping. Study and improvement measures for wet bags of orthopaedic exotic rental instruments[J]. Journal of Frontiers of Medicine, 2016.6(25):392-393
LI A H. Cause analysis and preventive measures of wet package in pulsating vacuum pressure steam sterilization[J]. China Med Device Inf, 2020, 26(13): 173-177.
LI Q B, LIN Q L, LIN B H. Analysis and countermeasures on orthopaedic sterilizing wet package of overweight closed rigid container box[J]. Chin J Dial Artif Organs, 2014, 25(3): 9-12.
ZHAO Y L, HU H, SHENG H. Observation on the failure mode and effect analysis effect of external instruments management and application in department of orthopaedics[J]. J Qilu Nurs, 2020, 26(10): 101-103.
WANG S Y, MA J L, ZHANG B S, et al. Comparison of the results of autoclave sterilization parameter monitoring and PCD monitoring[J]. Chin J Disinfect, 2020, 37(8): 573-575.
HAO S Q, DU H Y, LIU C, et al. Effect of loading density of different rigid containers on wet package of orthopedic instruments in the orthopedics department[J]. Mod Clin Nurs, 2017, 16(6): 26-29.
XU Y M, CHEN Y H, JIN C P, et al. Phenomena of wet pack after pulsing vacuum steam sterilization: analysis and countermeasures[J]. Med Forum, 2014, 18(13): 1659-1661.
CEN Q. Advances in management of wet pack in central sterile supply department[J]. J Nurs Sci, 2012, 27(11): 95-96.
CAI X J, LI T Y, CHENG P. Study on application effect of package improvement scheme of foreign medical devices[J]. Chin J Disinfect, 2018, 35(4): 305-306.