Wang Qun,Mo Junjun,Lv Bei,et al.Study on the cleaning and loading method of ventilator lines of different lengths using automatic spray washer[J].Sterile Supply,2022,01(03):105-109.
Wang Qun,Mo Junjun,Lv Bei,et al.Study on the cleaning and loading method of ventilator lines of different lengths using automatic spray washer[J].Sterile Supply,2022,01(03):105-109. DOI: 10.11910/j.issn.2791-2043.2022.3.05.
Objective,2,To explore a reliable and effective loading method for cleaning ventilator lines with an automatic spray washer so as to improve the pass rate of ventilator lines.,Methods,2,Ventilator lines of different lengths were randomly selected, 180 each for the control group and the experimental group. In the control group, no specific requirements were made on the front and rear position of the pressure jet port inserting into the ventilator lines, the hang length of the pipeline end and the fixation. In the experimental group, the front and rear positions of the pressure jet port were selected according to the length of the ventilator lines, with the 30 cm long pipeline inserted in the rear row, the 60 cm long pipeline in the middle row and the 120 cm long pipeline in the front row. Choose the same pipeline cleaning procedure for cleaning and disinfection. The cleaning effect is evaluated in two approaches: method of visual inspection with illuminated magnifier and STF cleaning test card.,Results,2,The pass rate of the visual inspection was 99.4% in the control group and 100% in the experimental group. The pass rate of the two groups was ,P,>,0.05, representing no statistical significance. The pass rate of the STF cleaning test card was 88.9% in the control group and 98.3% in the experimental group. The pass rate of the two groups was X2=13.423,P,<,0.05, which was statistically significant. The pass rate of ventilator lines in the experimental group was significantly better than that in the control group.,Conclusion,2,The washing and disinfecting machine is effective to clean ventilator lines, a reliable and effective loading method helps to improve the pass rate and easy to operate, so that the quality of line cleaning is ensured and the nosocomial infection is prevented.
Ventilator lineAutomatic spray washerLoadingSTF cleaning test cardPass rate
HUANG R H. Discussion on hospital medical fixed assets management[J]. Account Learn, 2018(35): 161-162.
WU W P, XIE X M, MIAO Q M. Comparative analysis of different cleaning methods for ventilator pipeline[J]. China Med Device Inf, 2019, 25(11): 180-182.
ZHAO H J, WANG L H, WANG Y C, et al. Observation on disinfection effect of ventilator tubing by two disinfection methods[J]. Chin J Disinfect, 2020, 37(7): 508-509, 512.
CHEN F, ZHAO B X, SHEN J, et al. Investigation on contamination and pathogen distribution on ventilator circuits in hospital[J]. Chin J Disinfect, 2016, 33(5): 431-434.
Ministry of Health of the People’s Republic of China. Regulation of disinfection technique in healthcare settings: WS/T367-2012[S]. Beijing, 2012.
LI L M. Analysis on the effect of three kinds of cleaning and disinfection methods of ventilator line[J]. J Henan Univ Med Sci, 2017, 36(3): 218-220.
ZHANG Z C, PAN F, GONG J, et al. Performance and application of ventilator-pipe cleaning and disinfection machines[J]. Chin J Nosocomiology, 2011, 21(16): 3426-3427.
FENG Xiulan, PENG Gangyi, eds. Guidelines for the construction and management of hospital sterile supply centres[M]. Guangdong Science and Technology Press, 2011,11.
LIN C R, CHEN K Z, JIANG Y Q, et al. Study on effect of sterilization for ventilator pipes and accessories by full automatic disinfection machines[J]. China Health Stand Manag, 2017, 8(27): 151-153.
CHEN S Z, BAI X L, JIN W D, et al. Manufacture of annular cleaning frame and its application in cleaning of pulp cavity expander[J]. Chin J Nurs, 2022, 57(10): 1273-1276.
CHAI X Y, SHAO C M, YU L, et al. Application of 8S management mode in the whole flow management of ventilator pipeline in Central Sterile Supply Department[J]. Chin Nurs Manag, 2021, 21(9): 1404-1408.
NEARY G,BLOM AW,SHIEL AI,et al.Development and biological evaluation of fluorophosphonate-modifie`d hydroxyapatite for orthopaedic applications[J].J Mater Sci Mater Med,2018,29(8):1-10.
0
Views
3
Downloads
0
CSCD
Publicity Resources
Related Articles
Related Author
Related Institution