Study on the Current Status of Standardized Initial-use Processing Technology and Management of New Stainless Steel Instruments in Central Sterile Supply Departments in Shaanxi Province
技术创新|Updated:2026-10-27
|
Study on the Current Status of Standardized Initial-use Processing Technology and Management of New Stainless Steel Instruments in Central Sterile Supply Departments in Shaanxi Province
Study on the Current Status of Standardized Initial-use Processing Technology and Management of New Stainless Steel Instruments in Central Sterile Supply Departments in Shaanxi Province
Sterile Supply2026年5卷第3期
Affiliations:
Central Sterile Supply Department, Shaanxi Provincial People’s Hospital, Xi’an 710068, Shaanxi Province
Li He, Zifeng Li, Qian Sun, 等. Study on the Current Status of Standardized Initial-use Processing Technology and Management of New Stainless Steel Instruments in Central Sterile Supply Departments in Shaanxi Province[J]. Sterile Supply, 2026,5(3):180-186.
He Li,Li Zifeng,Sun Qian,et al.Study on the Current Status of Standardized Initial-use Processing Technology and Management of New Stainless Steel Instruments in Central Sterile Supply Departments in Shaanxi Province[J].Sterile Supply,2026,05(03):180-186.
Li He, Zifeng Li, Qian Sun, 等. Study on the Current Status of Standardized Initial-use Processing Technology and Management of New Stainless Steel Instruments in Central Sterile Supply Departments in Shaanxi Province[J]. Sterile Supply, 2026,5(3):180-186.DOI: 10.11910/j.issn.2791-2043.2026.3.05.
He Li,Li Zifeng,Sun Qian,et al.Study on the Current Status of Standardized Initial-use Processing Technology and Management of New Stainless Steel Instruments in Central Sterile Supply Departments in Shaanxi Province[J].Sterile Supply,2026,05(03):180-186.DOI: 10.11910/j.issn.2791-2043.2026.3.05.
Study on the Current Status of Standardized Initial-use Processing Technology and Management of New Stainless Steel Instruments in Central Sterile Supply Departments in Shaanxi Province
To investigate the current implementation of initial-use pre-treatment of new stainless steel instruments in Central Sterile Supply Departments (CSSDs) of medical institutions in Shaanxi Province; identify deficiencies in current clinical management; establish a standardized pre-treatment procedure based on authoritative national and international standards; and develop feasible
tiered optimization measures to ensure sterilization safety and prolong instrument service life.
Methods
2
CSSD staff members from 107 medical institutions in the province qualified to process surgical stainless steel instruments were selected as study participants. A self-developed five-dimension questionnaire was used. Following expert evaluation of content validity and reliability testing in a pilot survey
an electronic questionnaire survey was conducted via the Wenjuanxing online platform
with quality control implemented throughout the entire process. A total of 107 valid questionnaires were collected
and descriptive statistics were performed using SPSS 26.0. Relevant national and international standards were systematically reviewed
and standardized procedures and tiered corrective strategies were developed based on the literature.
Results
2
Only 23.4% of institutions fully implemented the complete degreasing + passivation process
while 29.9% sterilized the instruments directly without pre-treatment. The utilization rates of dedicated degreasing agents and passivating agents were 32.7% and 8.4%
respectively. Only 33.6% of personnel were able to distinguish between degreasing and passivation
34.6% of institutions had established dedicated management procedures
and only 14.0% had incorporated pre-treatment into monthly quality control. Deficiencies existed across multiple aspects
including staff awareness
system development
operational implementation
consumable supplies
and quality control
resulting in a vicious cycle of management.
Conclusion
2
The level of standardization of pre-treatment for new stainless steel instruments in CSSDs in Shaanxi Province is relatively low. Medical institutions should implement corrective measures according to tiered priorities
strictly follow the standardized “degreasing-cleaning-passivation-lubrication and drying-sterilization” process
ensure adequate provision of dedicated reagents
improve management systems and tiered training
establish continuous quality control
and correct the misconception that “new instruments do not require pre-treatment”
thereby preventing instrument corrosion and staining at the source and ensuring the safety of aseptic clinical diagnosis and treatment.
关键词
Keywords
references
ZHANG X X , WANG H . Discussion on the standardized method of removing spots on the medical stainless steel surface and maintenance method [J ] . China Standardization , 2021 ( 12 ): 174 - 175 .
Specification for the management of yellow stains on instrument surfaces after sterilization [EB/OL ] . ( 2025-05-20 )[ 2026-08-25 ] .
Study on the effect of passivation treatment on reducing the rusting rate of stainless steel surgical instruments [J ] . China Medical Devices Information , 2024 , 30 ( 23 ): 173 - 175 .
YE H Y , YE Z M , WANG G H . Application of passivation in cleaning of loaned orthopedic instruments with lumen tube [J ] . Nursing of Integrated Traditional Chinese and Western Medicine , 2018 , 4 ( 6 ): 103 - 105 .
Specification for Reusable Surgical Instruments Management in Hospitals: T/NAHIEM 142—2025 [S ] .
Standard Guide for Care and Handling of Stainless Steel Surgical Instruments : ASTM F1744-96(2016) [S ] . ASTM International , 2016 .
ANSI/AAMI ST79 Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health Care Facilities [S ] .
State Administration for Market Regulation , National Standardization Administration . Surgical Instruments - Non-cutting, Articulated Instruments - General Requirements and Test Methods : GB/T 2766—2022 [S ] . Beijing : Standard Press of China , 2022 .
Training Textbook for Hospital Central Sterile Supply Department Personnel (2022 Edition) [M ] . Beijing : People’s Medical Publishing House , 2022 .
The trial reading is over, you can activate your VIP account to continue reading.
A Novel Swiss Cheese Model-Based Safety Chain Management System for Central Sterile Supply Department
Effects of Digital Closed-loop Traceability Management on Surgical Instrument Coordination Between the Central Sterile Supply Department and the Operating Room
Innovative Application of a Synergistic ATP Bioluminescence Assay–Video-in-Video Visualization–Traceability System for Closed-Loop Management of the Central Sterile Supply Department
Application Progress of Failure Mode and Effects Analysis in the Risk Management of Central Sterile Supply Departments
An AI-Assisted Traceability System for Risk Alerting and Efficiency Enhancement in the Central Sterile Supply Department: A Self-Controlled Study
相关作者
Min Li
Yunlei Li
Xiuhui Wu
Qiongrong Liu
Jing Che
Yuanyuan Song
Pengqin Gao
Jun Wang
相关机构
Jingmen Central Hospital Affiliated to Jingchu University of Technology
The People’s Hospital of Leshan
Central Sterile Supply Department, Xiangyang Central Hospital
Central Sterile Supply Department, Affiliated Hospital of Panzhihua University