Ling Fang, Qiang Zhang, Yu Liu, 等. Exploration of Most Difficult-to-Sterilize Location in Pulse Vacuum Sterilizers[J]. Sterile Supply, 2025,4(1):44-48.
Ling Fang,Qiang Zhang,Yu Liu,et al.Exploration of Most Difficult-to-Sterilize Location in Pulse Vacuum Sterilizers[J].Sterile Supply,2025,04(01):44-48.
Ling Fang, Qiang Zhang, Yu Liu, 等. Exploration of Most Difficult-to-Sterilize Location in Pulse Vacuum Sterilizers[J]. Sterile Supply, 2025,4(1):44-48. DOI: 10.11910/j.issn.2791-2043.2025.1.06.
Ling Fang,Qiang Zhang,Yu Liu,et al.Exploration of Most Difficult-to-Sterilize Location in Pulse Vacuum Sterilizers[J].Sterile Supply,2025,04(01):44-48. DOI: 10.11910/j.issn.2791-2043.2025.1.06.
Objective
2
The present study aims to explore the most difficult-to-sterilize location within the chamber of pulse vacuum sterilizers
providing a reference for the placement of biological monitoring packs during routine operation.
Method
2
Using temperature and pressure detectors
temperature testing was conducted on 16 identical models of pulse vacuum sterilizers in four Grade Ⅲ-A hospitals under different loading conditions
including no-load
instrument load
and dressing load. The location of the slowest heating point within the sterilizer chamber was determined.
Result
2
Statistics from 43 batches
totaling 258 data points show that the slowest heating point appeared most frequently at the lower level of the sterilizing rack
near the loading door
occurring 29 times
accounting for 67.44%; the location at the upper level of the sterilizing rack near the loading door occurred 14 times
accounting for 32.56%.
Conclusion
2
The most difficult-to-sterilize locations for this model of pulse vacuum sterilizer were near the loading side door
with the highest probability at the lower-level location near the door. Therefore
it is recommended to place the biological monitoring pack at the lower level of the sterilizing rack
near the loading side door
during routine biological monitoring.
NIU Z D , LU Y H , LIU W J , et al . Validation of sterilization effect of pulsant vacuum sterilizer [J ] . China Pharm , 2022 , 31 ( 22 ): 75 - 77 .
National Health Industry Standard of the People’s Republic of China . Central sterile supply department (CSSD) - Parts 1-3: Management Standards ( WS310.1-2016 ) (2016 Edition) [S ] . 2016 .
Jr YANG J , ROBBINS M , REILLY J , et al . The clinical effect of a rotator cuff retear: a meta-analysis of arthroscopic single-row and double-row repairs [J ] . Am J Sports Med , 2017 , 45 ( 3 ): 733 - 741 .
CHEN M , XU W , DONG Q R , et al . Outcomes of single-row versus double-row arthroscopic rotator cuff repair: a systematic review and meta-analysis of current evidence [J ] . Arthroscopy , 2013 , 29 ( 8 ): 1437 - 1449 .
DEHAAN A M , AXELRAD T W , KAYE E , et al . Does double-row rotator cuff repair improve functional outcome of patients compared with single-row technique?A systematic review [J ] . Am J Sports Med , 2012 , 40 ( 5 ): 1176 - 1185 .
General Administration of Quality Supervision , Inspection and Quarantine of the People’s Republic of China , Standardization Administration of China . Technical requirements for large steam sterilizers—Automatic type ( GB 8599—2008 ) [S ] . Beijing : General Administration of Quality Supervision, Inspection and Quarantine of the People’s Republic of China, Standardization Administration of China , 2008 .
GUAN L J , WEI J G , YANG L F . Study of sterilization parameters monitoring of external equipment during the sterilization process [J ] . Chongqing Medicine , 2021 , 50 ( 16 ): 2797 - 2799 .
China Pharmaceutical Association of Plant Engineering , Parenteral Drug Association (PDA) . PDA Technical Report No . 1 , Validation of Moist Heat Sterilization Processes : Cycle Design, Development, Qualification and Ongoing Control . 2007: Supplement Vol . 1 No. S-1.
SHEN J W , ZHANG Y X . Validation of heat distribution in autoclave [J ] . Pharm Eng Des , 2009 , 30 ( 6 ): 24 - 26 .
CHAI H R , YANG L L , ZHANG L . Study on testing sterilization performance of pulse vacuum sterilizer in hospital by temperature and pressure detector [J ] . J Hebei Med Univ , 2018 , 39 ( 6 ): 725 - 728 .
GU J H , ZHANG J Y , LU L X , et al . Application of high temperature and high pressure recorder in monitoring of vacuum pressure steam sterilizer [J ] . Chin J Nosocomiology , 2016 , 26 ( 21 ): 5022 - 5024 .