Xingling Deng at dxl129@163.com
纸质出版:2024-10-15
Scan QR Code
Jinyun Gao, Lijun Cai, Tingting Lin, 等. Comparison of cleaning effect of three methods on phacoemulsification silicone tube in ophthalmology[J]. 消毒供应, 2024,3(3):152-155.
Gao Jinyun,Cai Lijun,Lin Tingting,et al.Comparison of cleaning effect of three methods on phacoemulsification silicone tube in ophthalmology[J].Sterile Supply,2024,03(03):152-155.
Jinyun Gao, Lijun Cai, Tingting Lin, 等. Comparison of cleaning effect of three methods on phacoemulsification silicone tube in ophthalmology[J]. 消毒供应, 2024,3(3):152-155. DOI: 10.11910/j.issn.2791-2043.2024.3.07.
Gao Jinyun,Cai Lijun,Lin Tingting,et al.Comparison of cleaning effect of three methods on phacoemulsification silicone tube in ophthalmology[J].Sterile Supply,2024,03(03):152-155. DOI: 10.11910/j.issn.2791-2043.2024.3.07.
Objective
2
The present study aims to observe the cleaning effect of three methods on the phacoemulsification (phaco) silicone tube in ophthalmology.
Methods
2
A total of 142 phaco silicone tubes taken within 2 hours after surgery were divided in three groups randomly. Manual cleaning was applied in Group A (43 cases)
the method for cleaning and drying devices for batch cleaning of ophthalmic class A lumen was applied in Group B (46 cases)
and the method for cleaning with a negative-pressure vacuum cleaner was applied in Group C (53 cases). The cleaning time was recorded and the cleaning effect was evaluated by three methods
namely
adenosinetriphosphate (ATP) bioluminescence assay
magnifying glass with light
and air gun.
Results
2
The qualified rates of manual cleaning
the method for cleaning and drying devices for batch cleaning of ophthalmic class A lumen
and the method for cleaning with a negative-pressure vacuum cleaner were 95.3%
94.3%
and 95.7%
respectively
and the difference was not statistically significant (
P
>
0.05). The average cleaning duration was 76 seconds per tube
29.3 seconds per tube
and 69.46 seconds per tube
respectively.
Conclusion
2
All three cleaning methods are available in the clinic
of which the method for cleaning and drying devices for batch cleaning of ophthalmic class A lumen has the highest efficiency.
P.Z. Yang and X.Q. Fan , Ophthalmology . 9th ed . Beijing : People’s Medical Publishing House , 2018 .
Ministry of Health of the People’s Republic of China . Regulation of disinfection technique in healthcare settings : WS/T 367—2012 [S ] . Beijing : Standards Press of China , 2012 .
Y. Ma and J. He , “ Research progress of toxic anterior segment syndrome ,” International Eye Science , vol. 17 , pp. 669 - 672 , 2017 .
X.L. Deng , J.H. Huang , C.M. Du , et al , “ Quality evaluation of manual cleaning of Phaco silicone water pipe in ophthalmology ,” Chinese Journal of Disinfection , vol. 34 , pp. 1180 - 1181 , 2017 .
R. Shen , A.D. Gao , Z.H. Ma , et al , “ Study on the application of ATP bioluminescence technology in the cleaning quality monitoring of flexible endoscope ,” China Health Standard Management , vol. 13 , pp. 182 - 185 , 2022 .
Central sterile supply department(CSSD): Part 3: Surveillance standard for cleaning, disinfection and sterilization : WS 310.3—2016 [S ] . Beijing : Standards Press of China , 2017 .
K.X. Wang and Y.M. Zhu , “ Advances in the study on the method detecting the cleaning effect of endoscopic surgical instruments ,” China Medicine and Pharmacy , vol. 12 , pp. 52 - 54 , 2022 .
Y.M. Xiao , X.F. Liu , J.P. Zuo , et al , “ Influence of bedside pretreatment on cleaning effect of ophthalmic phacoemulsification surgical instruments ,” Nursing Practice and Research , vol. 19 , pp. 3264 - 3266 , 2022 .
G.F. Chen , G.Y. Fang and J.H. Yang , “ Clinical analysis of toxic reaction syndrome of anterior segment after cataract surgery ,” Modern Medicine and Health Research Electronic Journal , vol. 3 , pp. 16 - 17 , 2019 .
0
浏览量
68
Downloads
0
CSCD
关联资源
相关文章
相关作者
相关机构
苏公网安备 32010602011062号