Application and study of luminal sterilization challenge device in hydrogen peroxide cryogenic plasma in the sterilization effect[J]. 消毒供应, 2022,1(2):70-74.
Wang Yanhua,Zheng Mingxiang.Application and study of luminal sterilization challenge device in hydrogen peroxide cryogenic plasma in the sterilization effect[J].Sterile Supply,2022,01(02):70-74.
Application and study of luminal sterilization challenge device in hydrogen peroxide cryogenic plasma in the sterilization effect[J]. 消毒供应, 2022,1(2):70-74. DOI: 10.11910/j.issn.2791-2043.2022.2.07.
Wang Yanhua,Zheng Mingxiang.Application and study of luminal sterilization challenge device in hydrogen peroxide cryogenic plasma in the sterilization effect[J].Sterile Supply,2022,01(02):70-74. DOI: 10.11910/j.issn.2791-2043.2022.2.07.
Objective
2
To explore the lumen sterilization challenge device in three types of hydrogen peroxide low-temperature plasma sterilization effect further improve the low-temperature plasma sterilization device qualified rate.
Methods
2
In November 2021-May 2022
CSSD in our hydrogen peroxide plasma five card sterilization challenge device respectively in three types of low temperature plasma sterilization device exhaust vent positions
for different internal diameters
different lengths
different materials of medical metal and non-metal cavity instruments in three types of different cycle program instrument sterilization effect.
Results
2
The hydrogen peroxide plasma class 5 card sterilization challenge device is placed at the exhaust port of STERRAD NX sterilization device
with standard cycle and high sterilization qualification rate (
P
>
0.05) with internal diameter of 1mm; only compatible with the advanced circulation of the machine
with low sterilization qualification rate. Hydrogen peroxide plasma type 5 card sterilization challenge device is placed at the STERRAD 100 S sterilizer vent
Short cycles and long cycles for common metallic nonmetallic medical devices
High sterilization qualified rate
No sterilization effect (
P
>
0.05); For the inner diameter of 0.7 mm
Low short cycle sterilization rate of 500 mm
High qualified rate of long-cycle sterilization
The sterilization effect was statistically different (
P
<
0.01); The inner diameter is 1mm
A length of 850 mm single channel or no lumen soft mirror is only compatible with the advanced cycle of the machine
High qualified rate of sterilization. Hydrogen peroxide plasma type 5 card sterilization challenge device is placed at the STERRAD 100NX sterilization device outlet
the standard cycle for common metal non-metallic medical devices and internal diameter of 0.7 mm
length of 500 mm stainless steel medical lumen sterilization qualified rate; internal diameter of 1mm
length of 850 mm single channel or no cavity soft mirror can only be compatible with the soft mirror cycle of the machine
the sterilization qualified rate is also high.
Conclusion
2
STERRAD NX standard cycle sterilized common metal and non-metallic medical devices with high qualified rate; STERRAD 100S short cycle sterilized common metal and non-metallic medical devices with 0.7 mm and 500 mm Stainless steel medical lumen and internal diameter 1mm
length 850 mm
STERRAD 100NX standard cycle for common metal non-metallic medical instruments and internal diameter 0.7 mm
length 500 mm
and 500 mm
soft mirror cycle for internal diameter 1mm
length 850 mm single channel or without sterilization
sterilization qualified rate.
WANG D M , LEI Z H . Causes and modification of the cancellation of hydrogen peroxide low temperature plasma sterilization cycle [J]. Chin J Infect Control , 2017 , 16 ( 11 ): 1078 - 1079, 1088 .
LI Y , PAN J Y , ZHUANG Y , et al . Application of five kinds of chemical indicators in the monitoring of low-temperature plasma sterilization of hydrogen peroxide [J]. Electron J Clin Med Lit , 2019 , 6 ( 60 ): 177 - 178 .
HAN L . Application effects of PDCA cycle management in hydrogen peroxide low-temperature plasma sterilization cycle [J]. Med J Chin People’s Health , 2021 , 33 ( 21 ): 198 - 200 .
DU Y K , LIU Y Y . Maintenance and troubleshooting of hydrogen peroxide low-temperature plasma sterilizer [J]. China Med Device Inf , 2020 , 26 ( 15 ): 178 - 179 .
0
浏览量
358
Downloads
0
CSCD
关联资源
相关文章
相关作者
相关机构
苏公网安备 32010602011062号