Comparison of the first cleaning quality effect of two different cleaning methods for metal precision surgical instruments[J]. 消毒供应, 2022,1(2):55-58.
Liu Meifeng,Lin Suying,Zhu Hongmei,et al.Comparison of the first cleaning quality effect of two different cleaning methods for metal precision surgical instruments[J].Sterile Supply,2022,01(02):55-58.
Comparison of the first cleaning quality effect of two different cleaning methods for metal precision surgical instruments[J]. 消毒供应, 2022,1(2):55-58. DOI: 10.11910/j.issn.2791-2043.2022.2.04.
Liu Meifeng,Lin Suying,Zhu Hongmei,et al.Comparison of the first cleaning quality effect of two different cleaning methods for metal precision surgical instruments[J].Sterile Supply,2022,01(02):55-58. DOI: 10.11910/j.issn.2791-2043.2022.2.04.
Comparison of the first cleaning quality effect of two different cleaning methods for metal precision surgical instruments
To explore effective methods for the first cleaning of metal precision surgical instruments and evaluate the effect to improve the quality of cleaning.
Methods
2
The 184 newly purchased precision surgical instruments used in departments including Otolaryngology
Thoracic surgery and Orthopedics were selected and divided into two groups. 94 pieces in the control group were treated in the method of soaking in the alkaline detergent with a ratio of 1:100 and a temperature of 60-80 °C for 10 minutes+mechanical cleaning (alkaline detergent program); 90 pieces in the observation group were treated in the method of soaking in the alkaline detergent with a ratio of 1∶100 and a temperature of 60-80 °C for 10 minutes and then manual cleaning+mechanical cleaning (alkaline detergent program). After cleaning the instruments in both groups
the quality of the lumen instruments was checked by methods of visual inspection
illuminated magnifier
white gauze
and white strip. The sterilization packages were also opened again after autoclave sterilization to evaluate the oily waste degradation.
Results
2
The pass rates of visual inspection and illuminated magnifier
white gauze and white strip
visual inspection and illuminated magnifier after sterilization
and white gauze and white strip after sterilization in observation group were significantly higher than those of control group (100.00% vs. 94.68%
98.89% vs. 92.55%
98.89% vs. 90.43%
100.00% vs. 91.49%
respectively
P
<
0.05).
Conclusion
2
To ensure cleaning quality
metal precision surgical instruments should be correctly cleaned before the first use. The method of soaking in alkaline detergent for 10 minutes+manual cleaning+mechanical cleaning (alkaline detergent program) can significantly improve the cleaning quality.
关键词
Keywords
references
HUANG H , ZHANG Q . Investigation on usage and treatment of precision surgical instruments in hospitals [J]. Chin J Nurs , 2017 , 52 ( 10 ): 1238 - 1240 .
CHEN Xujun . [Industrial cleaning agents and cleaning technology] [M]. Beijing : Chemical Industry Press , 2005 .
LIU Yucun , LIANG Minghui . [Job Tutorial of Hospital Disinfection Supply Center] [M]. Beijing : People's Military Medical Press , 2013 .
ZHANG Qing , QIAN Liming . [Sterilization Supply Center Management and Technical Guide] [M]. Beijing : People's Medical Publishing House , 2021 : 109 .
LU Y J , OU S Q . Different cleaning methods of toothed medical devices and their effect evaluation [J]. China Mod Med , 2017 , 24 ( 15 ): 139 - 141 .
ZHU S J , YU L Y , ZHOU X , et al . [Analysis of surface cleanliness of new recycling operation instruments] [J]. Chin J Disinfect , 2013 , 30 ( 4 ): 317 - 318 .
OU S Q , ZHANG H H , CAO S J , et al . [Cleaning method and effect of medical equipment before first use] [J]. Chin J Disinfect , 2017 , 34 ( 10 ): 977 - 978 .
The trial reading is over, you can activate your VIP account to continue reading.