Application of homemade drying rack in simple respirator drying[J]. 消毒供应, 2022,1(1):32-35.
Yang Yueling,Qiu Fei,Shen Jinghua,et al.Application of homemade drying rack in simple respirator drying[J].Sterile Supply,2022,01(01):32-35.
Application of homemade drying rack in simple respirator drying[J]. 消毒供应, 2022,1(1):32-35. DOI: 10.11910/j.issn.2791-2043.2022.1.08.
Yang Yueling,Qiu Fei,Shen Jinghua,et al.Application of homemade drying rack in simple respirator drying[J].Sterile Supply,2022,01(01):32-35. DOI: 10.11910/j.issn.2791-2043.2022.1.08.
Objective
2
To analyze the drying effect of two groups of simple respirator drying methods and explore a better drying method.
Methods
2
120 simple respirators cleaned and disinfected in the disinfection supply room of a 3A hospital in January 2021 were selected and randomly divided into control group (60 drying cabinets) and observation group (60 drying racks+drying cabinets). The remaining water stains in the respirators were absorbed and weighed with 0.3 g cotton balls after 10 and 15 minutes of drying respectively. The results of 10 and 15 minutes of drying were compared between the two groups 15 minutes of drying and the number of cases of complete drying
two groups of drying methods
all respirators complete drying time.
Results
2
After drying for 10 minutes
the weight gain of the control group was 2.45%±20 g
the weight gain of cotton ball in the observation group was 0.62±19 g
0 in the control group and 7 in the observation group
P=0.004; After drying for 15 minutes
the weight gain of cotton balls in the control group was 2.14%±52 g
the weight gain of cotton ball in the observation group was 0.29±0.04 g
the number of completely dry cases in the control group was 0
and the number of dry cases in the observation group was 19
P
=0.00. The complete drying time of the control group was 95.72±07 minutes; The complete drying time of the observation group was 13.10±75 minutes
P=0.00
the difference was statistically significant (
P
<
0.05).
Conclusion
2
Using self-made drying frame can shorten the drying time of simple respirator
improve the qualified rate of respirator drying and the working efficiency of disinfection supply center.
Classification of medical devices [EB/OL]. 2017-8-31 . http://www.nmpa.gov.cn/WS04/CL2138/300389.html http://www.nmpa.gov.cn/WS04/CL2138/300389.html .
Deepak Jain , Carmen D’Ugard , Ana Aguilar , Teresa Del Moral , Eduardo Bancalari and Nelson Claure . Use of a Mechanical Ventilator with Respiratory Function Monitoring Provides More Consistent Ventilation during Simulated Neonatal Resuscitation [J]. Neonatology , 2020 , 117 ( 2 ): 151 - 158 .
CAI B L , MA L , DAI T , et al . Study on simple breathing balloons’ shelf-time after high level disinfection [J]. China Health Stand Manag , 2017 , 8 ( 17 ): 125 - 127 .
Y Gomo , J Eilevstjnn , K Holte , A Yeconia and HL Ersdal . Delivery of Positive End-Expiratory Pressure Using Self-Inflating Bags during Newborn Resuscitation Is Possible Despite Mask Leak [J]. Neonatology , 2020 , 117 ( 3 ): 1 - 8 .
Yon Hee Shim , Jong Seok Lee , Jung In Lee , Dong hun Choe and Kyu Dae Shim . Severe Hypotension Caused by Valve Malfunction in the Self-Inflating Bag-Valve Unit: A case report . Unit : A case report . 2004 , 46 ( 3 ): 360 - 362 .
Wei Zhang and Xia Yin . Sterilizing and drying rack of humidifying pot for respiratory apparatus . [P] CN200994930Y , 2006 .
WU X Y . Optimization of simple respirator disinfection supply process [J]. World Latest Med Inf , 2015 , 15 ( 94 ): 151 , 154 .
LU A W , WANG M , NIE Z Q , et al . Sterilizing effect and residual bacteria of four different methods for sterilizing simple breathing vesicles [J]. J South Med Univ , 2013 , 33 ( 7 ): 1067 - 1070 .
0
浏览量
151
Downloads
0
CSCD
关联资源
相关文章
相关作者
相关机构
苏公网安备 32010602011062号