Analysis of the use of temperature and pressure detectors in the validation of pressure steam sterilization parameters
According to the specification requirements of pressure steam sterilization equipment parameters such as temperature
time and pressure monitoring.
Methods
2
Using the corresponding pressure steam sterilization equipment according to temperature and pressure detector parameter validation of sterilization process
the key analysis compare the change of temperature
time.
Results
2
Ten of the pressure steam sterilization equipment 40 monitoring stations
the temperature of 35 monitoring sites and bring the sterilization equipment shown in temperature data
temperature success rate is 87.5%; 33 points at the time of consistent with own sterilization equipment shown in time data
time success rate of 82.5%.
Conclusion
2
The application of temperature and pressure detector to parameter validation of pressure steam sterilization equipment
can accurately for sterilization equipment operation monitoring
such as temperature
time
solved the use department of parameters in the process of using pressure steam sterilization equipment。