He Xiaoying.Application Progress of Failure Mode and Effects Analysis in the Risk Management of Central Sterile Supply Departments[J].Sterile Supply,2026,05(02):127-139.
DOI:
He Xiaoying.Application Progress of Failure Mode and Effects Analysis in the Risk Management of Central Sterile Supply Departments[J].Sterile Supply,2026,05(02):127-139.DOI: 10.11910/j.issn.2791-2043.2026.2.07.
Application Progress of Failure Mode and Effects Analysis in the Risk Management of Central Sterile Supply Departments
To review the evidence regarding the application of Failure Mode and Effects Analysis (FMEA) in the risk management of Central Sterile Supply Departments (CSSDs)
compare improvement priorities across different risk scenarios
and propose a practical tiered implementation plan.
Methods
2
A narrative review was conducted by searching CNKI
Wanfang Data
VIP
and PubMed. Citation tracking was performed using publications from the World Health Organization
patient safety resources of the U.S. Department of Veterans Affairs
and recent quality control documents for sterile supply services. The search covered all literature published up to June 30
2026. The application scenarios
failure modes
interventions
outcome measures
and limitations of the evidence were descriptively summarized.
Results
2
A total of 22 representative publications were included. FMEA has been applied to the management of loaned medical devices
endoscopes
packaging and sterilization
occupational protection
and integration with other management tools
and its application has gradually expanded to regional shared sterile supply services
information-based traceability systems
and 3D-printed patient-specific medical devices. Most studies reported reductions in process defects and nonconformities following intervention. However
the available evidence was still dominated by single-center before-and-after studies
with substantial heterogeneity in scoring systems
observation periods
and outcome definitions.
Conclusion
2
FMEA is well suited to the prospective identification of failure risks and the prioritization of improvement efforts in high-risk CSSD processes
but the Risk Priority Number (RPN) should not replace professional judgment. A localized approach that combines severity-triggered prioritization with RPN-based ranking
supported by information traceability
tiered implementation checklists
and multicenter prospective validation
may enhance the sustainability and reproducibility of risk management.
关键词
Keywords
references
WORLD HEALTH ORGANIZATION . Decontamination and reprocessing of medical devices for health-care facilities [M ] . Geneva : World Health Organization , 2016 .
DEROSIER J , STALHANDSKE E , BAGIAN J P , et al . Using health care failure mode and effect analysis TM : The VA national center for patient safety’s prospective risk analysis system [J ] . The Joint Commission Journal on Quality Improvement , 2002 , 28 ( 5 ): 248 - 267 .
SONG Z H , QIN Y H . Research progress on the application of failure mode and effects analysis in the risk management of central sterile supply departments [J ] . Chinese General Practice Nursing , 2023 , 21 ( 16 ): 2184 - 2186 .
YARMOHAMMADIAN M , REZAEI F , HAGHSHENAS A , et al . Revised risk priority number in failure mode and effects analysis model from the perspective of healthcare system [J ] . International Journal of Preventive Medicine , 2018 , 9 ( 1 ): 7 .
NIE S N , SONG D H , MENG Y T . A study on improving the reprocessing quality of loaner medical instruments in a central sterile supply department based on failure mode and effects analysis [J ] . Chinese Journal of Disinfection , 2023 , 40 ( 10 ): 784 - 787 .
GUO L , MO J , WU Q , et al . Meta-analysis assessing the contemporary state of disposal practices for loaned instruments in central sterile supply department [J ] . Chinese Journal of Disinfection , 2024 , 41 ( 8 ): 610 - 614 .
SHI Q . Application of failure mode and effects analysis in risk management for the handling of laparoscopic instruments in the CSSD [J ] . Medical Equipment , 2025 , 38 ( 6 ): 37 - 40 .
WANG X F , CHEN D F , LIU Y Q , et al . Application of failure mode and effects analysis in the whole-process management of rigid endoscopic instruments in a central sterile supply department [J ] . Chinese Journal of Disinfection , 2024 , 41 ( 8 ): 599 - 602 .
FAN P C , LIU H , LIU W Z . Application of failure mode and effects analysis in evaluating the cleaning and disinfection effectiveness of flexible endoscopes [J ] . China Journal of Endoscopy , 2025 , 31 ( 4 ): 1 - 7 .
ZHENG L L , GAN L , TAO J , et al . Application of FMEA and PDCA cycle in centralized management of flexible endoscopes in a central sterile supply department [J ] . China Medical Devices , 2025 , 40 ( 11 ): 92 - 96 .
QIU X D , WANG L J , YAN L Z , et al . Effect of a failure mode and effects analysis-based management model on the sterilization quality of laparoscopic instruments [J ] . Chinese Journal of Disinfection , 2024 , 41 ( 3 ): 219 - 221 .
ZHANG Y R , HU R X , CHEN Y H , et al . Improving pressure steam sterilization quality through healthcare failure mode and effects analysis: A pre-post intervention study in central sterile supply departments [J ] . Risk Management and Healthcare Policy , 2025 , 18 : 2313 - 2321 .
YI L Y , CHEN Y H , HU R X , et al . Application of healthcare failure mode and effect analysis in controlling surgical instrument packaging defects [J ] . Scientific Reports , 2022 , 12 : 19708 .
GUÉDON A C P , WAUBEN L S G L , VAN DER EIJK A C , et al . Where are my instruments? Hazards in delivery of surgical instruments [J ] . Surgical Endoscopy , 2016 , 30 ( 7 ): 2728 - 2735 .
DU H C , GUAN H X , LIANG J X , et al . Current status of sharps injuries in the central sterile supply department of a tertiary hospital and application of FMEA-based risk management [J ] . Guide of China Medicine , 2024 , 22 ( 3 ): 88 - 90 .
HU J , SUN X M , CUI K K . Effect of HFMEA combined with QCC quality-control measures on infection prevention and process quality in a central sterile supply department [J ] . Hainan Medical Journal , 2023 , 34 ( 17 ): 2580 - 2584 .
BEN HAMMAMIA R , EL HOUDA BEN FATMA N , LASMAR M , et al . Risk analysis applied to the sterilization process in a public health hospital center: use of the FMECA method [J ] . La Tunisie Médicale , 2025 , 103 ( 9 ): 1303 - 1309 .
ZHAN R W , SHEN Y H , ZHENG R Y , et al . Application of failure mode and effects analysis combined with an information traceability system in the reprocessing workflow of reusable dental instruments [J ] . Chinese Journal of Disinfection , 2024 , 41 ( 6 ): 457 - 459 .
HU Y L , YU J , YANG H , et al . The effects of quality traceability systems on shared medical community disinfection centers [J ] . Scientific Reports , 2025 , 15 : 16396 .
ZHENG J , WANG L , FANG Y H , et al . Analysis on the application of FMEA in ‘instrument and equipment surface cleaning and disinfection’ in hospitals based on standardization and cleaning and disinfection information system management [J ] . Frontiers in Public Health , 2024 , 12 : 1444721 .
LAN D P , ZHOU Z Q , YANG Y , et al . Influence of material, sterilization, and disinfection on the accuracy of three-dimensional printed surgical templates: An in vitro study [J ] . Clinical Oral Implants Research , 2025 , 36 ( 2 ): 191 - 201 .
National Institute of Hospital Administration , NHC . Quality Control Indicators for Central Sterile Supply Services (2024 Edition): GWYYH [2024 ] No. 185 [R ] . Beijing : National Institute of Hospital Administration, NHC , 2024 .
The trial reading is over, you can activate your VIP account to continue reading.
Effects of Digital Closed-loop Traceability Management on Surgical Instrument Coordination Between the Central Sterile Supply Department and the Operating Room
Innovative Application of a Synergistic ATP Bioluminescence Assay–Video-in-Video Visualization–Traceability System for Closed-Loop Management of the Central Sterile Supply Department
An AI-Assisted Traceability System for Risk Alerting and Efficiency Enhancement in the Central Sterile Supply Department: A Self-Controlled Study
Dual-Mode QR Code and NFC Management Improves Equipment Full Life-cycle Traceability Efficiency in Central Sterile Supply Department
Reusable Medical Device Processing in Central Sterile Supply Departments: A Survey of Secondary and Tertiary Hospitals in Xiangtan City, China
Related Author
Yuanyuan Song
Pengqin Gao
Jun Wang
Xia Wu
Dan Li
Min Zhu
Xiang Chen
Yan Zeng
Related Institution
The People’s Hospital of Leshan
Central Sterile Supply Department, Xiangyang Central Hospital
Jingmen Central Hospital
Shenzhen Traditional Chinese Medicine Hospital
Central Sterile Supply Department, Taihe Hospital, Shiyan City